(Created page with " =='''ORTEC Electronics Modules'''== * (1) ORTEC 710 Quad High Voltage Bias Supply (1 - 1000 V) * (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF (con...") |
|||
Line 4: | Line 4: | ||
* (1) ORTEC 710 Quad High Voltage Bias Supply (1 - 1000 V) | * (1) ORTEC 710 Quad High Voltage Bias Supply (1 - 1000 V) | ||
− | * (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF | + | * (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF (conversion gain 45 mV/MeV) |
− | (conversion gain 45 mV/MeV) | ||
− | * (2) ORTEC 142B Preamplifier for detector input capacitance 100 to 400 pF | + | * (2) ORTEC 142B Preamplifier for detector input capacitance 100 to 400 pF (conversion gain 20 mV/MeV) |
− | (conversion gain 20 mV/MeV) | ||
− | * (2) ORTEC Model 590A Amplifier and Timing Single-Channel Analyzer | + | * (2) ORTEC Model 590A Amplifier and Timing Single-Channel Analyzer (SCA) |
− | (SCA) | ||
* (2) ORTEC Model 570 Amplifier | * (2) ORTEC Model 570 Amplifier |
Revision as of 11:35, 17 January 2022
ORTEC Electronics Modules
- (1) ORTEC 710 Quad High Voltage Bias Supply (1 - 1000 V)
- (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF (conversion gain 45 mV/MeV)
- (2) ORTEC 142B Preamplifier for detector input capacitance 100 to 400 pF (conversion gain 20 mV/MeV)
- (2) ORTEC Model 590A Amplifier and Timing Single-Channel Analyzer (SCA)
- (2) ORTEC Model 570 Amplifier
==ORTEC ULTRA Detectors==
===(2) (BU-013-050-1000-S)===
- Ion-Implanted Silicon Charged Particles Detectors
- Ultra-thin entrance window (500 Å) for optimum energy resolution (FWHM, 𝛼 =13 keV, 𝛽 = 7 keV)
- B Mount
- Detector size of 50 mm2
- Depletion Depth (Range) of 1000 um for energies ≤ 500 keV
- Bias Voltage: +115 V
===(2) (BU-012-050-100)===
- Ion-Implanted Silicon Charged Particles Detectors
- Ultra-thin entrance window (500 Å) for optimum energy resolution (FWHM, 𝛼 =12 keV , 𝛽 = 6 keV)
- B Mount
- Detector size of 50 mm2
- Depletion Depth (Range) of 100 um for energies < 200 keV
- Bias Voltage: +50 V
===(2) (BU-012-050-500)===