Difference between revisions of "UITF Mott Detectors"
Jump to navigation
Jump to search
(8 intermediate revisions by the same user not shown) | |||
Line 4: | Line 4: | ||
* (1) ORTEC 710 Quad High Voltage Bias Supply (1 - 1000 V) [https://www.ortec-online.com/products/electronics/power-supplies-and-nuclear-instrument-module-nim-bins/710] | * (1) ORTEC 710 Quad High Voltage Bias Supply (1 - 1000 V) [https://www.ortec-online.com/products/electronics/power-supplies-and-nuclear-instrument-module-nim-bins/710] | ||
− | * (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF ( | + | * (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF (purchased 2007) (S/N 0715225357, 07152257) [https://www.ortec-online.com/products/electronics/preamplifiers/142a-b-c] |
− | * (2) ORTEC | + | * (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF (purchased 2003) (S/N 2225, 2228) [https://www.ortec-online.com/products/electronics/preamplifiers/142a-b-c] |
* (2) ORTEC Model 590A Amplifier and Timing Single-Channel Analyzer (SCA) [https://www.ortec-online.com/products/electronics/amplifiers/590a] | * (2) ORTEC Model 590A Amplifier and Timing Single-Channel Analyzer (SCA) [https://www.ortec-online.com/products/electronics/amplifiers/590a] | ||
Line 33: | Line 33: | ||
− | ==='''(2) (BU-013-050- | + | ==='''(2) (BU-013-050-500)'''=== |
* Ion-Implanted Silicon Charged Particles Detectors | * Ion-Implanted Silicon Charged Particles Detectors | ||
− | * Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=13 keV, 𝛽=7 keV) | + | * Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=13 keV , 𝛽=7 keV) |
* B Mount | * B Mount | ||
Line 43: | Line 43: | ||
* Detector size of 50 mm<sup>2</sup> | * Detector size of 50 mm<sup>2</sup> | ||
− | * Depletion Depth (Range) of | + | * Depletion Depth (Range) of 500 um for energies ≤ 300 keV |
+ | |||
+ | * Bias Voltage: +180 V | ||
− | * | + | * S/N: 22061675, 22061676 |
− | ==='''(2) (BU- | + | ==='''(2) (BU-013-050-1000-S)'''=== |
* Ion-Implanted Silicon Charged Particles Detectors | * Ion-Implanted Silicon Charged Particles Detectors | ||
− | * Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼= | + | * Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=13 keV, 𝛽=7 keV) |
* B Mount | * B Mount | ||
Line 58: | Line 60: | ||
* Detector size of 50 mm<sup>2</sup> | * Detector size of 50 mm<sup>2</sup> | ||
− | * Depletion Depth (Range) of | + | * Depletion Depth (Range) of 1000 um for energies ≤ 500 keV |
+ | |||
+ | * Bias Voltage: +115 V | ||
− | * | + | * Serial No: 40-015A11, 40-015A12 |
− | ==='''(2) (BU- | + | ==='''(2) (BU-012-050-100)'''=== |
* Ion-Implanted Silicon Charged Particles Detectors | * Ion-Implanted Silicon Charged Particles Detectors | ||
− | * Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼= | + | * Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=12 keV , 𝛽=6 keV) |
* B Mount | * B Mount | ||
Line 73: | Line 77: | ||
* Detector size of 50 mm<sup>2</sup> | * Detector size of 50 mm<sup>2</sup> | ||
− | * Depletion Depth (Range) of | + | * Depletion Depth (Range) of 100 um for energies ≤ 100 keV |
+ | |||
+ | * Bias Voltage: +50 V | ||
− | * | + | * Serial No: 38-130Z15, 38-130Z16 |
Latest revision as of 13:01, 11 July 2022
ORTEC Electronics Modules
- (1) ORTEC 710 Quad High Voltage Bias Supply (1 - 1000 V) [1]
- (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF (purchased 2007) (S/N 0715225357, 07152257) [2]
- (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF (purchased 2003) (S/N 2225, 2228) [3]
- (2) ORTEC Model 590A Amplifier and Timing Single-Channel Analyzer (SCA) [4]
- (2) ORTEC Model 570 Amplifier [5]
Single-Channel Analyzer (SCA)
590A Amplifier and Timing Single-Channel Analyzer [6]
- SCA Out:
- Standard NIM of +5 V, 500 ns wide if the amplified signal passes the threshold of the Lower Level discriminator
- Connect to Scaler
- Amp Out:
- Energy output (Amplified: x5 to x1250)
- Connect to fADC
ORTEC ULTRA Detectors
ULTRA Ion Implanted Silicon Charged Particle Radiation Detectors [7]
(2) (BU-013-050-500)
- Ion-Implanted Silicon Charged Particles Detectors
- Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=13 keV , 𝛽=7 keV)
- B Mount
- Detector size of 50 mm2
- Depletion Depth (Range) of 500 um for energies ≤ 300 keV
- Bias Voltage: +180 V
- S/N: 22061675, 22061676
(2) (BU-013-050-1000-S)
- Ion-Implanted Silicon Charged Particles Detectors
- Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=13 keV, 𝛽=7 keV)
- B Mount
- Detector size of 50 mm2
- Depletion Depth (Range) of 1000 um for energies ≤ 500 keV
- Bias Voltage: +115 V
- Serial No: 40-015A11, 40-015A12
(2) (BU-012-050-100)
- Ion-Implanted Silicon Charged Particles Detectors
- Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=12 keV , 𝛽=6 keV)
- B Mount
- Detector size of 50 mm2
- Depletion Depth (Range) of 100 um for energies ≤ 100 keV
- Bias Voltage: +50 V
- Serial No: 38-130Z15, 38-130Z16