Difference between revisions of "Detectors"

From Ciswikidb
Jump to navigation Jump to search
 
(79 intermediate revisions by the same user not shown)
Line 1: Line 1:
==Detectors==
+
=='''Detectors'''==
  
* The detectors are described in the following POLOG Entry:
+
* The detectors are described in the following POLOG Entries:
  
http://opweb.acc.jlab.org/CSUEApps/elog02/elog_item.php?elog_id=1542515
 
  
https://cebaf.jlab.org/elog/entry/1542515
+
https://logbooks/entry/1542515
 +
 
 +
https://logbooks/entry/1542519
 +
 
 +
https://logbooks/entry/1542529
 +
 
 +
 
 +
 
 +
=='''Data Sheets'''==
  
 
  - The Acrylic Light Guide is 0.125 inch x 1 inch wide x 2 inch long (data sheet: [[media:Light_Guide.pdf]]).
 
  - The Acrylic Light Guide is 0.125 inch x 1 inch wide x 2 inch long (data sheet: [[media:Light_Guide.pdf]]).
  
  - The Plastic Scintillator for the DE detector is an EJ-212 machine finish of the size 1 mm x 1 inch x 1 inch (data sheet: [[media:EJ-212.pdf]]).
+
  - The Plastic Scintillator for the DE detector is an EJ-212 machine finish of the size 1 mm x 1 inch x 1 inch  
 +
(data sheet: [[media:EJ-212.pdf]]).
 +
 
 +
- The Plastic Scintillator Rod for the E detector is an EJ-200 (data sheet: [[media:EJ-200.pdf]]) of the size
 +
3 inch diameter x 2.5 inch long painted with EJ-510 (data sheet: [[media:EJ-510.pdf]]) with one end clear.
 +
 
 +
- The 1-inch PMT is R6427 ([[media:PMT_R6427.pdf]]) in an H7415 Assembly.
 +
 
 +
- The 3-inch PMT is R6091 ([[media:PMT_R6091.pdf]]) in an H6559 Assembly.
 +
 
 +
 
 +
 
 +
=='''Old Logic Setup'''==
 +
 
 +
[[file:Mott_Logic.jpg||600px|]]
 +
 
 +
[[file:TEK0377.jpg||400px|]]
 +
 
 +
 
 +
 
 +
* ΔE Detector PMT HV:
 +
 
 +
{| class="wikitable"
 +
|-
 +
| '''LEFT'''
 +
| '''-1000 V'''
 +
|-
 +
| '''RIGHT'''
 +
| '''-1000 V'''
 +
|-
 +
| '''UP'''
 +
| '''-1000 V''' 
 +
|-
 +
| '''DOWN'''
 +
| '''-1080 V''' 
 +
|-
 +
|}
 +
 
 +
 
 +
* E Detector PMT HV:
 +
 
 +
{| class="wikitable"
 +
|-
 +
| '''LEFT'''
 +
| '''-1125 V'''
 +
|-
 +
| '''RIGHT'''
 +
| '''-1120 V'''
 +
|-
 +
| '''UP'''
 +
| '''-1140 V''' 
 +
|-
 +
| '''DOWN'''
 +
| '''-1160 V''' 
 +
|-
 +
|}
 +
 
 +
-----
 +
 
 +
 
 +
 
 +
* ΔE Detector 792 Dual Delay Module ([[media:792ds.pdf]]):
 +
 
 +
{| class="wikitable"
 +
|-
 +
| '''LEFT'''
 +
| '''0.5+4+16+32 ns'''
 +
|-
 +
| '''RIGHT'''
 +
| '''0.5+1+2+16+32 ns'''
 +
|-
 +
| '''UP'''
 +
| '''1+2+16+32 ns '''
 +
|-
 +
| '''DOWN'''
 +
| '''0.5+2+16+32 ns'''
 +
|-
 +
|}
 +
 
 +
-----
 +
 
 +
 
 +
 
 +
* ΔE Detector 705 Octal Discriminator Veto Thresholds ([[media:705ds.pdf]]):
 +
 
 +
{| class="wikitable"
 +
|-
 +
| '''LEFT'''
 +
| '''-185 mV'''
 +
|-
 +
| '''RIGHT'''
 +
| '''-189 mV'''
 +
|-
 +
| '''UP'''
 +
| '''-185 mV'''
 +
|-
 +
| '''DOWN'''
 +
| '''-186 mV'''
 +
|-
 +
|}
 +
 
 +
 
 +
* ΔE Detector 715 Constant Fraction Timing Discriminator Thresholds ([[media:715ds.pdf]]):
 +
 
 +
'''Shaping Delay = 2 ns'''
 +
 
 +
{| class="wikitable"
 +
|-
 +
| '''LEFT'''
 +
| '''-25 mV'''
 +
|-
 +
| '''RIGHT'''
 +
| '''-29 mV'''
 +
|-
 +
| '''UP'''
 +
| '''-27 mV'''
 +
|-
 +
| '''DOWN'''
 +
| '''-42 mV'''
 +
|-
 +
|}
 +
 
 +
 
 +
 
 +
* E Detector 715 Constant Fraction Timing Discriminator Thresholds ([[media:715ds.pdf]]):
 +
 
 +
'''Shaping Delay = 4 ns'''
 +
 
 +
{| class="wikitable"
 +
|-
 +
| '''LEFT'''
 +
| '''-25 mV'''
 +
|-
 +
| '''RIGHT'''
 +
| '''-25 mV'''
 +
|-
 +
| '''UP'''
 +
| '''-30 mV'''
 +
|-
 +
| '''DOWN'''
 +
| '''-30 mV'''
 +
|-
 +
|}
 +
 
 +
 
 +
-----
 +
 
 +
=='''New Logic Setup'''==
 +
 
 +
The above is what we had in December 2013 and earlier. On Jan 13, 2014:
 +
 
 +
1- Adjusted HV such that signals of ΔE and E are '''-200 mV'''. FADC range = 0 to -500 mV, 0 to 2<sup>12</sup> = 4096.
 +
2- Set E thresholds = '''-25 mV'''.
 +
3- Set ΔE thresholds = '''-25 mV'''.
 +
4- Set the Veto thresholds to '''-450 mV'''.
 +
 
 +
 
 +
 
 +
* ΔE Detector PMT New HV:
 +
 
 +
{| class="wikitable"
 +
|-
 +
| '''LEFT'''
 +
| '''-1300 V'''
 +
|-
 +
| '''RIGHT'''
 +
| '''-1320 V'''
 +
|-
 +
| '''UP'''
 +
| '''-1230 V''' 
 +
|-
 +
| '''DOWN'''
 +
| '''-1320 V''' 
 +
|-
 +
|}
 +
 
 +
 
 +
 
 +
 
 +
* E Detector PMT New HV:
 +
 
 +
{| class="wikitable"
 +
|-
 +
| '''LEFT'''
 +
| '''-1200 V'''
 +
|-
 +
| '''RIGHT'''
 +
| '''-1185 V'''
 +
|-
 +
| '''UP'''
 +
| '''-1220 V''' 
 +
|-
 +
| '''DOWN'''
 +
| '''-1225 V''' 
 +
|-
 +
|}
 +
 
 +
 
 +
 
 +
 
 +
* ΔE Detector New Delays:
 +
 
 +
{| class="wikitable"
 +
|-
 +
| '''LEFT'''
 +
| '''16+32 ns'''
 +
|-
 +
| '''RIGHT'''
 +
| '''16+32 ns'''
 +
|-
 +
| '''UP'''
 +
| '''16+32 ns'''
 +
|-
 +
| '''DOWN'''
 +
| '''16+32 ns'''
 +
|-
 +
|}
 +
 
 +
 
 +
 
 +
* E and ΔE signal triggered from the E signal with a trigger threshold of -50 mV. PMT HVs are adjusted such that all signals are -200 mV.
 +
 
 +
 
 +
[[file:EDE_Raw_Signals_LEFT.jpg|left|300px|]]
 +
[[file:EDE_Raw_Signals_RIGHT.jpg|right|300px|]]
 +
[[file:EDE_Raw_Signals_UP.jpg|left|300px|]]
 +
[[file:EDE_Raw_Signals_DOWN.jpg|right|300px|]]
 +
 
 +
 
 +
[[file:ADCdEs_7298.gif||500px|]]
 +
 
 +
 
 +
[[file:ADCEs_7298.gif||500px|]]
 +
 
 +
 
 +
 
 +
-----
 +
 
  
- The Plastic Scintillator Rod for the E detector is an EJ-200 (data sheet: [[media:EJ-200.pdf]]) of the size 3 inch diameter x 2.5 inch long painted with EJ-510 (data sheet: [[media:EJ-510.pdf]]) with one end clear.
+
* E trigger gate (30 ns) and ΔE trigger gate (5 ns) triggered from E signal (with ΔE delays listed ) and the E-ΔE Coincidence signal.
  
- The 1-inch PMT is R6427 ([[media:PMT_R6427.pdf]]) in an H7415 Assembly (Figure 4).
 
  
- The 3-inch PMT is R6091 ([[media:PMT_R6427.pdf]) in an H6559 Assembly (Figures 5 and 6).
+
[[file:EDE_Disc_Signals_LEFT.jpg|left|300px|]]
 +
[[file:EDE_Disc_Signals_RIGHT.jpg|right|300px|]]
 +
[[file:EDE_Disc_Signals_UP.jpg|left|300px|]]
 +
[[file:EDE_Disc_Signals_DOWN.jpg|right|300px|]]

Latest revision as of 07:51, 25 April 2014

Detectors

  • The detectors are described in the following POLOG Entries:


https://logbooks/entry/1542515

https://logbooks/entry/1542519

https://logbooks/entry/1542529


Data Sheets

- The Acrylic Light Guide is 0.125 inch x 1 inch wide x 2 inch long (data sheet: media:Light_Guide.pdf).
- The Plastic Scintillator for the DE detector is an EJ-212 machine finish of the size 1 mm x 1 inch x 1 inch 
(data sheet: media:EJ-212.pdf).
- The Plastic Scintillator Rod for the E detector is an EJ-200 (data sheet: media:EJ-200.pdf) of the size 
3 inch diameter x 2.5 inch long painted with EJ-510 (data sheet: media:EJ-510.pdf) with one end clear.
- The 1-inch PMT is R6427 (media:PMT_R6427.pdf) in an H7415 Assembly.
- The 3-inch PMT is R6091 (media:PMT_R6091.pdf) in an H6559 Assembly.


Old Logic Setup

Mott Logic.jpg

TEK0377.jpg


  • ΔE Detector PMT HV:
LEFT -1000 V
RIGHT -1000 V
UP -1000 V
DOWN -1080 V


  • E Detector PMT HV:
LEFT -1125 V
RIGHT -1120 V
UP -1140 V
DOWN -1160 V


LEFT 0.5+4+16+32 ns
RIGHT 0.5+1+2+16+32 ns
UP 1+2+16+32 ns
DOWN 0.5+2+16+32 ns


LEFT -185 mV
RIGHT -189 mV
UP -185 mV
DOWN -186 mV


  • ΔE Detector 715 Constant Fraction Timing Discriminator Thresholds (media:715ds.pdf):

Shaping Delay = 2 ns

LEFT -25 mV
RIGHT -29 mV
UP -27 mV
DOWN -42 mV


  • E Detector 715 Constant Fraction Timing Discriminator Thresholds (media:715ds.pdf):

Shaping Delay = 4 ns

LEFT -25 mV
RIGHT -25 mV
UP -30 mV
DOWN -30 mV



New Logic Setup

The above is what we had in December 2013 and earlier. On Jan 13, 2014:

1- Adjusted HV such that signals of ΔE and E are -200 mV. FADC range = 0 to -500 mV, 0 to 212 = 4096. 
2- Set E thresholds = -25 mV.
3- Set ΔE thresholds = -25 mV.
4- Set the Veto thresholds to -450 mV.


  • ΔE Detector PMT New HV:
LEFT -1300 V
RIGHT -1320 V
UP -1230 V
DOWN -1320 V



  • E Detector PMT New HV:
LEFT -1200 V
RIGHT -1185 V
UP -1220 V
DOWN -1225 V



  • ΔE Detector New Delays:
LEFT 16+32 ns
RIGHT 16+32 ns
UP 16+32 ns
DOWN 16+32 ns


  • E and ΔE signal triggered from the E signal with a trigger threshold of -50 mV. PMT HVs are adjusted such that all signals are -200 mV.


EDE Raw Signals LEFT.jpg
EDE Raw Signals RIGHT.jpg
EDE Raw Signals UP.jpg
EDE Raw Signals DOWN.jpg


ADCdEs 7298.gif


ADCEs 7298.gif




  • E trigger gate (30 ns) and ΔE trigger gate (5 ns) triggered from E signal (with ΔE delays listed ) and the E-ΔE Coincidence signal.


EDE Disc Signals LEFT.jpg
EDE Disc Signals RIGHT.jpg
EDE Disc Signals UP.jpg
EDE Disc Signals DOWN.jpg