Difference between revisions of "Detectors"
Jump to navigation
Jump to search
Line 156: | Line 156: | ||
* Step 1: E and ΔE signal triggered from the E signal with a trigger threshold of -25 mV. | * Step 1: E and ΔE signal triggered from the E signal with a trigger threshold of -25 mV. | ||
+ | |||
[[file:EDE_Raw_Signals_LEFT.jpg|left|300px|]] | [[file:EDE_Raw_Signals_LEFT.jpg|left|300px|]] | ||
[[file:EDE_Raw_Signals_RIGHT.jpg|right|300px|]] | [[file:EDE_Raw_Signals_RIGHT.jpg|right|300px|]] | ||
Line 162: | Line 163: | ||
* Step 2: E trigger gate (30 ns) and ΔE trigger gate (5 ns) triggered from E signal (with ΔE delays listed above). | * Step 2: E trigger gate (30 ns) and ΔE trigger gate (5 ns) triggered from E signal (with ΔE delays listed above). | ||
+ | |||
[[file:EDE_Disc_Signals_LEFT.jpg|left|300px|]] | [[file:EDE_Disc_Signals_LEFT.jpg|left|300px|]] | ||
[[file:EDE_Disc_Signals_RIGHT.jpg|right|300px|]] | [[file:EDE_Disc_Signals_RIGHT.jpg|right|300px|]] | ||
[[file:EDE_Disc_Signals_UP.jpg|left|300px|]] | [[file:EDE_Disc_Signals_UP.jpg|left|300px|]] | ||
[[file:EDE_Disc_Signals_DOWN.jpg|right|300px|]] | [[file:EDE_Disc_Signals_DOWN.jpg|right|300px|]] |
Revision as of 10:23, 7 January 2014
Detectors
- The detectors are described in the following POLOG Entry:
http://opweb.acc.jlab.org/CSUEApps/elog02/elog_item.php?elog_id=1542515
https://cebaf.jlab.org/elog/entry/1542515
Data Sheets
- The Acrylic Light Guide is 0.125 inch x 1 inch wide x 2 inch long (data sheet: media:Light_Guide.pdf).
- The Plastic Scintillator for the DE detector is an EJ-212 machine finish of the size 1 mm x 1 inch x 1 inch (data sheet: media:EJ-212.pdf).
- The Plastic Scintillator Rod for the E detector is an EJ-200 (data sheet: media:EJ-200.pdf) of the size 3 inch diameter x 2.5 inch long painted with EJ-510 (data sheet: media:EJ-510.pdf) with one end clear.
- The 1-inch PMT is R6427 (media:PMT_R6427.pdf) in an H7415 Assembly.
- The 3-inch PMT is R6091 (media:PMT_R6091.pdf) in an H6559 Assembly.
Logic Setup
- ΔE Detector PMT HV:
LEFT | -1000 V |
RIGHT | -1000 V |
UP | -1000 V |
DOWN | -1080 V |
- E Detector PMT HV:
LEFT | -1125 V |
RIGHT | -1120 V |
UP | -1140 V |
DOWN | -1160 V |
- ΔE Detector 792 Dual Delay Module (media:792ds.pdf):
LEFT | 1+2+8 ns |
RIGHT | 4+8 ns |
UP | 0.5+4+8 ns |
DOWN | 1+4+8 ns |
- ΔE Detector 705 Octal Discriminator Thresholds (media:705ds.pdf):
LEFT | -185 mV |
RIGHT | -189 mV |
UP | -185 mV |
DOWN | -186 mV |
- ΔE Detector 715 Constant Fraction Timing Discriminator Thresholds (media:715ds.pdf):
Shaping Delay = 2 ns
LEFT | -25 mV |
RIGHT | -29 mV |
UP | -27 mV |
DOWN | -42 mV |
- E Detector 715 Constant Fraction Timing Discriminator Thresholds (media:715ds.pdf):
Shaping Delay = 4 ns
LEFT | -25 mV |
RIGHT | -25 mV |
UP | -30 mV |
DOWN | -30 mV |
Mott Electronics Logic Setup Procedure
- Step 1: E and ΔE signal triggered from the E signal with a trigger threshold of -25 mV.
- Step 2: E trigger gate (30 ns) and ΔE trigger gate (5 ns) triggered from E signal (with ΔE delays listed above).