Bubble Chamber Beam Test May 2018
Jump to navigation
Jump to search
Bubble Chamber Turn On
- Fill with natural C3F8 – test bubble chamber systems operation
- With beam on bubble chamber radiator:
- How does CCD camera perform under beam-on conditions?
- Count rates on bubble chamber. Do we get single or multiple bubbles from Bremsstrahlung beam exposure?
- Measure gamma ray beam spatial profile as reflected by bubble distribution. Is collimator effective in defining the gamma-ray beam?
- Test the bubble chamber laser shutter (gumby shutter)
- Background measurements:
- Measure beam off environmental background in chamber-injector area
- Measure beam on background by looking outside fiducial volume
- Measure background with beam to Faraday Cup in CEBAF beamline (about two meters from chamber)
- Measure neutron events in chamber. Neutron radiation detectors in injector region will indicate if any neutrons are generated
Bubble Chamber Beam Test
Run Plan:
- Check if Bubble Chamber is working properly:
- Start with K.E. of 5.2 MeV
- Determine the initial operational pressure and temperature
- Measure bubble rate. The expected rate is 1 bubble per sec at 10 µA
- Measure rate vs beam current (0.01, 0.1 and 1.0 µA)
- Measure the number of bubbles for 30 minutes per beam current
Beam Kinetic Energy (MeV) | Beam Current (µA) | Time (hour) |
4.7 | 2 | 5 |
4.8 | 2 | 5 |
4.9 | 1 | 5 |
5.0 | 1 | 5 |
5.1 | 0.8 | 5 |
5.2 | 0.4 | 5 |
5.3 | 0.2 | 5 |
- Presentations
- MCC 8 am presentation