UITF Mott Detectors

From Ciswikidb
Revision as of 12:36, 4 May 2022 by Suleiman (talk | contribs)
Jump to navigation Jump to search

ORTEC Electronics Modules

  • (1) ORTEC 710 Quad High Voltage Bias Supply (1 - 1000 V) [1]
  • (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF (conversion gain 45 mV/MeV) [2]
  • (2) ORTEC 142B Preamplifier for detector input capacitance 100 to 400 pF (conversion gain 20 mV/MeV) [3]
  • (2) ORTEC Model 590A Amplifier and Timing Single-Channel Analyzer (SCA) [4]
  • (2) ORTEC Model 570 Amplifier [5]



Single-Channel Analyzer (SCA)

590A Amplifier and Timing Single-Channel Analyzer [6]

  • SCA Out:
    • Standard NIM of +5 V, 500 ns wide if the amplified signal passes the threshold of the Lower Level discriminator
    • Connect to Scaler
  • Amp Out:
    • Energy output (Amplified: x5 to x1250)
    • Connect to fADC



ORTEC ULTRA Detectors

ULTRA Ion Implanted Silicon Charged Particle Radiation Detectors [7]


(2) (BU-013-050-1000-S)

  • Ion-Implanted Silicon Charged Particles Detectors
  • Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=13 keV, 𝛽=7 keV)
  • B Mount
  • Detector size of 50 mm2
  • Depletion Depth (Range) of 1000 um for energies ≤ 500 keV
  • Bias Voltage: +115 V


(2) (BU-012-050-100)

  • Ion-Implanted Silicon Charged Particles Detectors
  • Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=12 keV , 𝛽=6 keV)
  • B Mount
  • Detector size of 50 mm2
  • Depletion Depth (Range) of 100 um for energies ≤ 100 keV
  • Bias Voltage: +50 V


(2) (BU-013-050-500)

  • Ion-Implanted Silicon Charged Particles Detectors
  • Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=13 keV , 𝛽=7 keV)
  • B Mount
  • Detector size of 50 mm2
  • Depletion Depth (Range) of 500 um for energies ≤ 300 keV
  • Bias Voltage: 180 V