UITF Mott Detectors

From Ciswikidb
Revision as of 13:23, 17 January 2022 by Suleiman (talk | contribs)
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

ORTEC Electronics Modules

  • (1) ORTEC 710 Quad High Voltage Bias Supply (1 - 1000 V) [1]
  • (2) ORTEC 142A Preamplifier for detector input capacitance 0 to 100 pF (conversion gain 45 mV/MeV) [2]
  • (2) ORTEC 142B Preamplifier for detector input capacitance 100 to 400 pF (conversion gain 20 mV/MeV) [3]
  • (2) ORTEC Model 590A Amplifier and Timing Single-Channel Analyzer (SCA) [4]
  • (2) ORTEC Model 570 Amplifier [5]



Single-Channel Analyzer (SCA)

[6]

  • SCA Out:
    • Standard NIM of +5 V, 500 ns wide if the amplified signal passes the threshold of the Lower Level discriminator
    • Connect to Scaler
  • Amp Out:
    • Energy output (Amplified: x5 to x1250)
    • Connect to fADC



ORTEC ULTRA Detectors

ULTRA Ion Implanted Silicon Charged Particle Radiation Detectors [7]


(2) (BU-013-050-1000-S)

  • Ion-Implanted Silicon Charged Particles Detectors
  • Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=13 keV, 𝛽=7 keV)
  • B Mount
  • Detector size of 50 mm2
  • Depletion Depth (Range) of 1000 um for energies ≤ 500 keV
  • Bias Voltage: +115 V


(2) (BU-012-050-100)

  • Ion-Implanted Silicon Charged Particles Detectors
  • Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=12 keV , 𝛽=6 keV)
  • B Mount
  • Detector size of 50 mm2
  • Depletion Depth (Range) of 100 um for energies ≤ 100 keV
  • Bias Voltage: +50 V


(2) (BU-013-050-500)

  • Ion-Implanted Silicon Charged Particles Detectors
  • Ultra-thin entrance window (500Å) for optimum energy resolution (FWHM, 𝛼=13 keV , 𝛽=7 keV)
  • B Mount
  • Detector size of 50 mm2
  • Depletion Depth (Range) of 500 um for energies ≤ 300 keV
  • Bias Voltage: ? V