UITF Notes

From Ciswikidb
Revision as of 09:46, 19 August 2022 by Bruker (talk | contribs)
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

MeV beam time

  • Measure MeV dispersion with BPMs => Done, need to analyze
  • TM beam to 703 to see RF drift (full shift)
  • Verify optics setup for HKDL
  • Repeat buncher study, booster crested, 8 MeV, multiple scans per point

keV beam time

  • LVQE scan at different voltages. Expect same result, provided different laser powers at same voltage give same result (avoid SCL).
    • Results are different, don't know if related to instrumentation (battery resistance etc.) or physics. If it's physics:
      • CST: Does thermal energy affect collection efficiency?
      • Ion current? probably too low to matter
      • Try adjustable bias voltage source (Keithley)
  • HVQE scan at different voltages. Expect different result.
  • High current to FC2 for 12+ hours, check if drift is gone
  • Look at mains-harmonic BPM signals as a function of parameters (buncher etc.)

Beam line modifications