Difference between revisions of "Recent Changes and Current Issues"

From clas12-run
Jump to navigation Jump to search
(Replaced content with " moved to run group wiki (RG-K)")
Tag: Replaced
 
(83 intermediate revisions by 4 users not shown)
Line 1: Line 1:
* '''Do not spend more than 15-20 min trying to fix a problem'''. With any issue, contact On-Call Experts or RC.
 
  
* <font color=BROWN>'''NEW: when starting a new run after a successful end, do not go from Prestart, but Abort, Download, Prestart, and Go.'''</font>
+
moved to run group wiki (RG-K)
 
 
* Especially after the HWP change, check the SLM and FCUP '''charge asymmetry''' (Q-Asym on Beamline Overview); the absolute values should be less than 0.1%. If they are out of range, contact the Beamline Expert.
 
 
 
* NOTE: the beam position on BPM 2C21A is not essential during production; the beam orbit lock is on BPM 2H01.
 
 
 
* '''RICH''':
 
:&loz; '''New RICH recovery protocol''': see the "{{#switchtablink:Shift_Expert|Shift Expert}}" tab the up-to-date information. '''Avoid unnecessary recoveries'''.
 
:&loz; RICH Sector 1 has three missing tiles; this is known, and no action is required from the shift.
 
 
 
* '''BMT/FMT/SVT:''' If new hot channels appear, '''call the MVT/SVT/MM expert Yuri Gotra''' (757-541-7539). These channels need to be addressed by the expert.
 
:&loz; '''<font color=red> Known Problem</font>''': Due to BMT SEC3 L1 subsequent trips, its HV was set to "0 V" so its occupancy plot is blank except from some noisy channels and thus looks different than the current reference run (r18663) distribution.
 
 
 
* '''Tagger magnet procedure:''' If the tagger magnet needs to be turned off, tell the operator to use '''only the script for degaussing, NEVER do "by hand"!'''
 
 
 
* '''Change of target in the hall:''' In case of questions from MCC about the necessary controlled access, refer to the logbook entry #: '''[https://logbooks.jlab.org/entry/4196113 4196113]''' for the access requirements. We do need escorted access when we had been at >=90nA (on the CuSn targets).
 
 
 
* '''Controlled access to the hall:'''
 
** At 50-60 nA beam current, we ask MCC to stop the beam down, wait for the beacon to stop flashing (15-20 minutes), and then do not need an escort to go in.
 
** At 95 nA beam current, the beacon will not stop flashing anytime soon, so we need escorted access, viz the ARM. In that case, we can keep the beam up until the ARM arrives at the counting house.
 

Latest revision as of 19:11, 17 January 2024

moved to run group wiki (RG-K)