Daily RC Meetings at 13:00 in CH Room 200C, and on ZOOM: URL Link (Meeting ID: 1611984219; Passcode: RGD-RC23), except on Wednesdays, which is on ZOOM ONLY @ 15:30.
Do NOT change FSD settings without approval from RC or beamline expert!
Line 27: | Line 27: | ||
== General Instructions == | == General Instructions == | ||
− | * The | + | * The main lights in the Hall and the Forward Carriage lights must be kept off during data taking. If these lights are switched on during an access, they '''should be switched off when leaving the Hall'''. |
** Refer to these photos: ON: [https://logbooks.jlab.org/entry/3825712], OFF: [https://logbooks.jlab.org/entry/3825731] | ** Refer to these photos: ON: [https://logbooks.jlab.org/entry/3825712], OFF: [https://logbooks.jlab.org/entry/3825731] | ||
** We have a new dome light switch outside of the hall in the labyrinth. The lights can be turned off without entering the Hall. See logbook entry: https://logbooks.jlab.org/entry/4204460 | ** We have a new dome light switch outside of the hall in the labyrinth. The lights can be turned off without entering the Hall. See logbook entry: https://logbooks.jlab.org/entry/4204460 | ||
** Note that when the dome lights are switched off, they cannot be turned back on immediately because they require 10-15 min to cool down. | ** Note that when the dome lights are switched off, they cannot be turned back on immediately because they require 10-15 min to cool down. | ||
− | * Beamline | + | * Beamline |
− | ** '''Do not run over XX minutes above YY nA with a 5-pass beam without the | + | ** '''Do not run over XX minutes above YY nA with a 5-pass beam without the beam blocker IN.''' The beam blocker must be IN for long-running at high currents for 5-pass operations. |
** If there is any concern about beam stability, ask MCC if orbit locks are on (they should be). | ** If there is any concern about beam stability, ask MCC if orbit locks are on (they should be). | ||
** Document any beam condition change and send scaler GUIs to HBLOG. | ** Document any beam condition change and send scaler GUIs to HBLOG. | ||
Line 39: | Line 39: | ||
* For beam tuning and Møller runs | * For beam tuning and Møller runs | ||
− | + | ** Make sure the SVT/MVT are set as follows: '''SVT: LV ON; HV OFF''', and '''MVT: LV ON; HV ON in SAFE Mode.''' | |
− | + | ** When '''turning MVT OFF''', do not press "BMT HV OFF" from HV->Overview->Actions, but instead '''select BMT HV ON - Safe Voltage'''. The same applies to FMT '''select FMT HV ON - Safe Voltage'''. | |
− | + | ** When turning MVT ON, go to HV->Overview->Actions, and select BMT (/FMT) "HV ON - Full Voltage". | |
− | + | ** DO NOT start runs in SAFE Mode for Production/Calibration/Various-Studies. | |
− | + | ** Turn <b> DC HV </b> off only for beam tuning; if no beam is available or when the beam is stable, keep them on even if you are not taking data. | |
− | * | + | * To access the Hall |
− | + | ** If you need to enter Hall B, alert Denny Insley (757-897-9060, cell/text: 757-810-9576) and the Run Coordinator (757-575-7540). | |
− | + | ** Check in with the shift personnel before entering the hall. | |
|} | |} | ||
− | {| {{TableStyle1}} | + | {| {{TableStyle1}} |
+ | |||
== <font color=Navy> '''DAQ Operation & related General Instructions''' </font> == | == <font color=Navy> '''DAQ Operation & related General Instructions''' </font> == | ||
| valign=top width=50% | | | valign=top width=50% | |
Revision as of 12:02, 26 February 2024
Run Coordinator: Hayk Hakyoban (757) 575-7540
PDL: Nathan Baltzell (9-) 757 876-1789; Office: x5902
Important Notes
- In case of medical or fire emergency, call 9-911 from a LANDLINE phone in the Counting House and inform the Crew Chief and RC.
- In case of evacuation of the Counting Room (e.g., due to a fire alarm), grab the shift expert phone and leave the room through the "EXIT" door leading to the parking lot. Do not leave through the hallway! Call the RC once outside so the latter will call MCC to turn the beam OFF to Hall B and inform other Hall B staff members.
- If you have symptoms or a positive COVID test, do NOT come to your shift - instead, inform the PDL and the RC and JLab medical services immediately to arrange replacements.
Monday, March 10 - Monday, Dec. 4
|
General Instructions
|
|
|
Reference Values
FSD Thresholds from RG-D
- Upstream: 2 kHz
- Midstream: 2 kHz
- Downstream: 100 kHz
- BOM: 50 kHz
- 5 ms dwell time
Harp Scan References from RG-D
- 2H01 [3]; x= -1.4 mm; y= 0.0 mm
- Tagger (2C24A) [4]: x= -2.0 mm; y= +0.6 mm
- 2C21A [5]: x= -0.4 mm; y= +0.6 mm (less important; values form RG-C)