Cosmic Run Notes:
 
<hr>
 
'''Tuesday March 1st'''
 
6:39pm : CDC turned on and run started
 
DAQ configuration issues
 
automated system running (image)
 
  
'''Wed:'''
+
== Objectives: ==
Temperature found to be less controlled, temps on edge or outside model training. Temperature fixed to 299.25
 
"ideal_gcf": 0.15292712143476878
 
 
 
Naomi not happy with the controls.  rec_scale changed to 15 min (15) under the assumption there was a build up of
 
David finds and fixes bug in cdc_tool.py which caused the setting to take more than the original rec_scale of 5 min.  Once fixed rec_scale reset to 5
 
new: "ideal_gcf": 0.15240693525146817
 
 
 
'''Thursday:'''
 
Naomi not happy with the recommended HV given the pressure.
 
 
 
control system reinitialized. "ideal_gcf": 0.15202469799843948
 
 
 
Naomi notices the default pressure is wrong and corrects it
 
 
 
"ideal_gcf": 0.1523301284531037
 
 
 
11:45am Thomas finds bug in initialization not respecting the fixed values.  Bug fixed. "ideal_gcf": 0.1482670653025542
 
 
 
12:21pm Torri starts a new run post bug fix (Run 90716)
 
 
 
1pm-ish beam current returning N/A forcing system to report bad epics.  This crashes the model and program.  Thomas routes around this N/A as beam current isn't being used and protects the model from this.  It is also logged in the DB now
 
 
 
2:05pm EPICS has died (more N/A) across the board.  Program no longer crashes but, as expected, the CDC is no longer under control of the AI
 
 
 
epics down between 13:13:13 -> 14:17:14 the pressure didn't change during this period so no harm done to the run. discussed a fail safe mechanism
 
 
 
'''Friday'''
 
pretty big pressure change. HV following as expected.
 
[[File:thur_fri_overnight.png]]
 
 
 
8:43am Thomas adds the fail safe with no interruption to running. System configuration now stands at:
 
{"model_name": "/gluonraid2/AIEC/models/d_cdc_modelneeds/model_pickle4_scilearn0242.pkl", "poll_time": 1, "rec_scale": 5, "control_window": [2110.0, 2150.0], "epics_trailing_window": 15, "ideal_gcf": 0.1482670653025542, "default_hvbi": -9.0, "require_human": 0, "control_mask": "BRB,BRT", "debug": 0, "default_temp": -299.25, "default_pressure": -100.022,"baseline_V":2130,"fail_safe_timeout":90}
 
 
 
Objectives:  
 
 
* Test drive the AIEC software
 
* Test drive the AIEC software
 
* Demonstrate that the CDC gain is stable
 
* Demonstrate that the CDC gain is stable
  
Schedule:
+
== Schedule: ==
 
* 2022-02-25 Fri. - Test fully automated CDC HV control
 
* 2022-02-25 Fri. - Test fully automated CDC HV control
 
** Notify Beni and Mark Stevens of intent to run test on Fri. throughout the day with low voltages (~20V-30V)
 
** Notify Beni and Mark Stevens of intent to run test on Fri. throughout the day with low voltages (~20V-30V)
Line 73: Line 33:
  
  
Strategy:
+
== Strategy: ==
 
* Designate 2130V as the nominal HV for cosmics, to raise more hits above the noise, but keep 20V headroom for the AI.
 
* Designate 2130V as the nominal HV for cosmics, to raise more hits above the noise, but keep 20V headroom for the AI.
 
* Tell the AI that the HVB current is constant at 9uA so that it can operate within the range of its trained data
 
* Tell the AI that the HVB current is constant at 9uA so that it can operate within the range of its trained data
Line 80: Line 40:
  
  
Needs:
+
== Needs: ==
 
* AIEC software to read epics and wait for a stable period before calculating the gain and setting the HV
 
* AIEC software to read epics and wait for a stable period before calculating the gain and setting the HV
 
** What determines "stable"?
 
** What determines "stable"?
Line 95: Line 55:
 
* <s>Code to set whole quadrants to same value</s>([https://github.com/JeffersonLab/AIEC/blob/main/utilities/cdc_tool.py cdc_tool.py])
 
* <s>Code to set whole quadrants to same value</s>([https://github.com/JeffersonLab/AIEC/blob/main/utilities/cdc_tool.py cdc_tool.py])
  
Questions:
+
== Questions: ==
 
* <s>Do we need to force a B field in the reconstruction sw (wait for Simon to answer)</s>
 
* <s>Do we need to force a B field in the reconstruction sw (wait for Simon to answer)</s>
 
** ''Simon reported progress on straight track reconstruction (email to David and Naomi Feb. 24th)''
 
** ''Simon reported progress on straight track reconstruction (email to David and Naomi Feb. 24th)''
Line 104: Line 64:
 
** ''March 1-14, 2022''
 
** ''March 1-14, 2022''
 
* Can we write something (epics data) to the evio file whenever the HV is changed?  Or increase the epics inclusion rate?
 
* Can we write something (epics data) to the evio file whenever the HV is changed?  Or increase the epics inclusion rate?
 +
 +
<hr>
 +
<hr>
 +
 +
== Cosmic Run Notes: ==
 +
<hr>
 +
=== 2022-03-01 ===
 +
'''Tuesday March 1st'''
 +
6:39pm : CDC turned on and run started
 +
DAQ configuration issues
 +
automated system running (image)
 +
 +
=== 2022-03-02 ===
 +
'''Wed:'''
 +
Temperature found to be less controlled, temps on edge or outside model training. Temperature fixed to 299.25
 +
"ideal_gcf": 0.15292712143476878
 +
 +
Naomi not happy with the controls.  rec_scale changed to 15 min (15) under the assumption there was a build up of
 +
David finds and fixes bug in cdc_tool.py which caused the setting to take more than the original rec_scale of 5 min.  Once fixed rec_scale reset to 5
 +
new: "ideal_gcf": 0.15240693525146817
 +
 +
=== 2022-03-03 ===
 +
'''Thursday:'''
 +
Naomi not happy with the recommended HV given the pressure.
 +
 +
control system reinitialized. "ideal_gcf": 0.15202469799843948
 +
 +
Naomi notices the default pressure is wrong and corrects it
 +
 +
"ideal_gcf": 0.1523301284531037
 +
 +
11:45am Thomas finds bug in initialization not respecting the fixed values.  Bug fixed. "ideal_gcf": 0.1482670653025542
 +
 +
12:21pm Torri starts a new run post bug fix (Run 90716)
 +
 +
1pm-ish beam current returning N/A forcing system to report bad epics.  This crashes the model and program.  Thomas routes around this N/A as beam current isn't being used and protects the model from this.  It is also logged in the DB now
 +
 +
2:05pm EPICS has died (more N/A) across the board.  Program no longer crashes but, as expected, the CDC is no longer under control of the AI
 +
 +
epics down between 13:13:13 -> 14:17:14 the pressure didn't change during this period so no harm done to the run. discussed a fail safe mechanism
 +
 +
=== 2022-03-04 ===
 +
'''Friday'''
 +
pretty big pressure change. HV following as expected.
 +
[[File:thur_fri_overnight.png]]
 +
 +
8:43am Thomas adds the fail safe with no interruption to running. System configuration now stands at:
 +
{"model_name": "/gluonraid2/AIEC/models/d_cdc_modelneeds/model_pickle4_scilearn0242.pkl", "poll_time": 1, "rec_scale": 5, "control_window": [2110.0, 2150.0], "epics_trailing_window": 15, "ideal_gcf": 0.1482670653025542, "default_hvbi": -9.0, "require_human": 0, "control_mask": "BRB,BRT", "debug": 0, "default_temp": -299.25, "default_pressure": -100.022,"baseline_V":2130,"fail_safe_timeout":90}

Revision as of 19:44, 4 March 2022

Objectives:

  • Test drive the AIEC software
  • Demonstrate that the CDC gain is stable

Schedule:

  • 2022-02-25 Fri. - Test fully automated CDC HV control
    • Notify Beni and Mark Stevens of intent to run test on Fri. throughout the day with low voltages (~20V-30V)
    • Modify control script to subtract 2100V from suggested setting before applying
    • Create cronjob under hdops account to run script every 5min.
    • Monitor system occasionally throughout the day so that it runs for 4-6hrs
  • 2022-03-01 Tue. - Start system
    • Notify halld@jlab.org that the test is starting and provide contact info.
    • Verify control script no longer subtracts 2100V from suggested setting
    • Confirm CDC gas is flowing and bubbler is correct
    • Confirm CDC alcohol level is good
    • Confirm BCAL chiller is on
    • Turn on all CDC HV channels
    • Turn on all BCAL HV,LV channels
    • Start DAQ system and confirm that cosmic trigger rate is correct (~100Hz?)
      • TODO: Get correct DAQ configuration details from Sergey
    • Set cronjob under hdops account to run script every 2min. and watch for a few cycles
    • Check CDC occupancy in RootSpy
      • TODO: Confirm that the occupancy histogram will work with cosmic events
      • TODO: Add additional monitoring plots to RootSpy
    • Increase the frequency of writing EPICS into the datastream.
    • If all seems to be functioning, reset the crontab to run the script once per hour
    • Make HDLOG logbook entry
    • Monitor progress regularly for 2 weeks. Start new runs if needed.
  • 2022-03-14 Mon. - End test
    • Put all CDC HV setpoints back to 2125V
    • Change the EPICS frequency back to normal.


Strategy:

  • Designate 2130V as the nominal HV for cosmics, to raise more hits above the noise, but keep 20V headroom for the AI.
  • Tell the AI that the HVB current is constant at 9uA so that it can operate within the range of its trained data
  • Use 1V steps and 1h runs
    • Starting new run every hour for 2 weeks may not be feasible. Devise scheme to break up long runs?


Needs:

  • AIEC software to read epics and wait for a stable period before calculating the gain and setting the HV
    • What determines "stable"?
    • Who?
  • Method for scaling the HVB current to what it would be at the nominal HV before running the AI - will not be needed for cosmics but will be needed for CPP.
    • Is this something that could be tested during cosmics run? (yes=implement, no=defer)
  • Script to project dE/dx for 4-10 GeV tracks, fit it and calculate the gain factor Not needed for cosmics.
    • Is this for needed for cosmics? (monitoring?)
  • CDC_occupancy script needs to be adapted for cosmics if we intend to use rootspy.
  • CDC_amp plugin adapted to make histograms for masked wires. Done.
  • Code to set whole quadrants to same value(cdc_tool.py)

Questions:

  • Do we need to force a B field in the reconstruction sw (wait for Simon to answer)
    • Simon reported progress on straight track reconstruction (email to David and Naomi Feb. 24th)
  • Full DAQ or standalone DAQ?
    • Ask Sergey for configuration with only CDC, BCAL and writing only to one file at a time.
    • For run 80239 the 19GB files contained 1/3 M triggers each. This file size is ok.
  • When are we doing this?
    • March 1-14, 2022
  • Can we write something (epics data) to the evio file whenever the HV is changed? Or increase the epics inclusion rate?


Cosmic Run Notes:


2022-03-01

Tuesday March 1st 6:39pm : CDC turned on and run started DAQ configuration issues automated system running (image)

2022-03-02

Wed: Temperature found to be less controlled, temps on edge or outside model training. Temperature fixed to 299.25 "ideal_gcf": 0.15292712143476878

Naomi not happy with the controls. rec_scale changed to 15 min (15) under the assumption there was a build up of David finds and fixes bug in cdc_tool.py which caused the setting to take more than the original rec_scale of 5 min. Once fixed rec_scale reset to 5 new: "ideal_gcf": 0.15240693525146817

2022-03-03

Thursday: Naomi not happy with the recommended HV given the pressure.

control system reinitialized. "ideal_gcf": 0.15202469799843948

Naomi notices the default pressure is wrong and corrects it

"ideal_gcf": 0.1523301284531037

11:45am Thomas finds bug in initialization not respecting the fixed values. Bug fixed. "ideal_gcf": 0.1482670653025542

12:21pm Torri starts a new run post bug fix (Run 90716)

1pm-ish beam current returning N/A forcing system to report bad epics. This crashes the model and program. Thomas routes around this N/A as beam current isn't being used and protects the model from this. It is also logged in the DB now

2:05pm EPICS has died (more N/A) across the board. Program no longer crashes but, as expected, the CDC is no longer under control of the AI

epics down between 13:13:13 -> 14:17:14 the pressure didn't change during this period so no harm done to the run. discussed a fail safe mechanism

2022-03-04

Friday pretty big pressure change. HV following as expected. Thur fri overnight.png

8:43am Thomas adds the fail safe with no interruption to running. System configuration now stands at: {"model_name": "/gluonraid2/AIEC/models/d_cdc_modelneeds/model_pickle4_scilearn0242.pkl", "poll_time": 1, "rec_scale": 5, "control_window": [2110.0, 2150.0], "epics_trailing_window": 15, "ideal_gcf": 0.1482670653025542, "default_hvbi": -9.0, "require_human": 0, "control_mask": "BRB,BRT", "debug": 0, "default_temp": -299.25, "default_pressure": -100.022,"baseline_V":2130,"fail_safe_timeout":90}