Overview
(see Tritium DAQ page)
Trigger Checkout Plan (Experts Only)
In general, we need one hall access if everything works fine.
Step 1
- Use remote scopes with inputs S0 (from FIFO), S2 (from FIFO), GC (from FIFO), S0&S2 (from Logic Unit) and S0||S2 (from Logic Unit)
- Verify that the S0 and S2 signals are as they should be (see picture (needs to be added!!!))
- Verify that the S0&&S2 signal is within the GC signal with the desired time distance (see picture (needs to be added!!!))
- Verify that the S0||S2 signal is within the GC signal with the desired time distance. Take also into account the extra cable delay from the S0|S2 logic unit to the (S0||S2)&&GC logic unit (see picture (needs to be added!!!))
- NOTE: The S0&&S2 and S0||S2 times could be also estimated from the S0 and S2 signals with the known delays from the trigger schematics if there is no remote scope available for the RHRS.
- If necessary go to the hall and change individual trigger delays!
- In parallel: Check S0, S2 and GC timings with Tong's and Jason's scripts.
Step 2
- Go to the hall and plug in the trigger signals and S2 on the 4-channel scopes.
- Check the individual trigger delays with respect to each and other and S2 (leads the time for S0&S2 and S0&S2&GC). Keep in mind to take also into account different cable length of the triggers from the logic unit to the TS input.
- If necessary go to the hall and modify the cables.
- In parallel: Check trigger time distributions with Shujie's script.