Difference between revisions of "Data Acquisition System and Triggers"
Jump to navigation
Jump to search
Line 26: | Line 26: | ||
== [https://wiki.jlab.org/tegwiki/index.php/Howto_access_remote_scopes Remote Scope Access] == | == [https://wiki.jlab.org/tegwiki/index.php/Howto_access_remote_scopes Remote Scope Access] == | ||
+ | |||
+ | |||
+ | == FADC Check == | ||
+ | 1.Modify FADC readout list to switch to raw sample mode | ||
+ | |||
+ | 2.Decode a run with FADC decoder: | ||
+ | gotritium; cd scripts/FADC_decode; | ||
+ | ./FADCcom | ||
+ | 3.check fadc_data.root. Any under/overflow? Is signal 15 samples away to the window beginning? |
Revision as of 23:09, 11 December 2017
Back to Installation and Commissioning
Overview
(see Tritium DAQ page)
Trigger Checkout Plan (Experts Only)
In general, we need one hall access if everything works fine.
Step 1
- Use remote scopes with inputs S0 (from FIFO), S2 (from FIFO), GC (from FIFO), S0&S2 (from Logic Unit) and S0||S2 (from Logic Unit)
- Verify that the S0 and S2 signals are as they should be (see picture (needs to be added!!!))
- Verify that the S0&&S2 signal is within the GC signal with the desired time distance (see picture (needs to be added!!!))
- Verify that the S0||S2 signal is within the GC signal with the desired time distance. Take also into account the extra cable delay from the S0|S2 logic unit to the (S0||S2)&&GC logic unit (see picture (needs to be added!!!))
- NOTE: The S0&&S2 and S0||S2 times could be also estimated from the S0 and S2 signals with the known delays from the trigger schematics if there is no remote scope available for the RHRS.
- If necessary go to the hall and change individual trigger delays!
- In parallel: Check S0, S2 and GC timings with Tong's and Jason's scripts.
Step 2
- Go to the hall and plug in the trigger signals and S2 on the 4-channel scopes.
- Check the individual trigger delays with respect to each and other and S2 (leads the time for S0&S2 and S0&S2&GC). Keep in mind to take also into account different cable length of the triggers from the logic unit to the TS input.
- If necessary go to the hall and modify the cables.
- In parallel: Check trigger time distributions with Shujie's script.
Remote Scope Access
FADC Check
1.Modify FADC readout list to switch to raw sample mode
2.Decode a run with FADC decoder:
gotritium; cd scripts/FADC_decode; ./FADCcom
3.check fadc_data.root. Any under/overflow? Is signal 15 samples away to the window beginning?