Difference between revisions of "Kerr - July 29, 2016"

From Ciswikidb
Jump to navigation Jump to search
 
(5 intermediate revisions by the same user not shown)
Line 9: Line 9:
  
 
'''Excel Sheets for MMF data taken'''
 
'''Excel Sheets for MMF data taken'''
First, Data taken with original setup. Stage untouched since the 28th. Before stage was realigned in an attempt to be parallel with foil
+
 
 +
 
 +
---First, Data taken with original setup. Stage untouched since the 28th. Before stage was realigned in an attempt to be parallel with foil
 
*two measurements were taken
 
*two measurements were taken
 
*Center was approximately 10.75 mm (on micrometer)
 
*Center was approximately 10.75 mm (on micrometer)
Line 16: Line 18:
 
* at 7.0 mm (on micrometer)  
 
* at 7.0 mm (on micrometer)  
 
** [[media:preadj7.0mm.xlsx]]
 
** [[media:preadj7.0mm.xlsx]]
 +
 +
--- '''adjustments made on linear stage in an attempt to gain parallel relation to foil, We moved legs of table to be parallel with solenoid.'''
 +
 +
Note: Center at 12.8 mm
 +
* far right 17.3 mm (on micrometer)
 +
** [[media:farright17.3mm.xlsx]]
 +
** This is not pretty^
 +
* right center at 15.4 mm (on Micrometer)
 +
** [[media:rightcenter15.4mm.xlsx]]
 +
** Also not pretty^
 +
* at 12.8 mm (on Micrometer) w/ lens... Note: I did not finish this measurement or foolishly did not save it...
 +
** [[media:12.8mmwlens.xlsx]]
 +
* at 12.8 mm (on Micrometer) w/out lens... this was a test to see if the lens was the reason for the "dip"... As soon as we saw a dip we agreed the lens was not affecting signal therefore we didn't finish the measurement
 +
** [[media:12.8mmwoutlens.xlsx]]
 +
 +
 +
* With mirror in place of foil at center 12.8 mm
 +
** [[media:12.8mmmirrorbackground.xlsx]]
 +
* 25 amp increments (as opposed to 50 amp inc.) at 12.8 mm (on micrometer)
 +
** [[media:12.8mm25inccenter.xlsx]]
 +
* Stability check at 12.8 mm (on Micrometer) with constant I of 200 amps over span of 16 min
 +
** [[media:stability200amps.xlsx]]

Latest revision as of 23:39, 29 July 2016

Magnetic Measurement Facility readings of pucks, Steel () and Molybdenum --> media:LDGT01 Puck Measurements for MOKE 16-7-28.xlsx

This graph reads 6.33 Gauss/1 Amp. Note this is only the magnetic field at the surface of the Steel () puck.


Important INFO to note

  • foil surface diameter across is 11 mm +-.5 mm
  • Laser dot diameter through polarizer to foil surface is 2mm +-.5 mm

Excel Sheets for MMF data taken


---First, Data taken with original setup. Stage untouched since the 28th. Before stage was realigned in an attempt to be parallel with foil

--- adjustments made on linear stage in an attempt to gain parallel relation to foil, We moved legs of table to be parallel with solenoid.

Note: Center at 12.8 mm

  • far right 17.3 mm (on micrometer)
  • right center at 15.4 mm (on Micrometer)
  • at 12.8 mm (on Micrometer) w/ lens... Note: I did not finish this measurement or foolishly did not save it...
  • at 12.8 mm (on Micrometer) w/out lens... this was a test to see if the lens was the reason for the "dip"... As soon as we saw a dip we agreed the lens was not affecting signal therefore we didn't finish the measurement