Difference between revisions of "Mott Run1 Target Thickness Study"
Line 1: | Line 1: | ||
;[[CEBAF_MeV_Mott_Polarimeter]] | ;[[CEBAF_MeV_Mott_Polarimeter]] | ||
− | Summary data: rates, thicknesses, asym [[Media:summary25Nov2015.xls]] , [[Media:summary25Nov2015. | + | Summary data: rates, thicknesses, asym [[Media:summary25Nov2015.xls]] , [[Media:summary25Nov2015.txt]] |
==Mott Run1 Target Thickness Study -- Best Data== | ==Mott Run1 Target Thickness Study -- Best Data== |
Latest revision as of 11:53, 25 November 2015
Summary data: rates, thicknesses, asym Media:summary25Nov2015.xls , Media:summary25Nov2015.txt
Mott Run1 Target Thickness Study -- Best Data
Provided in summary tables in link to .ods document below.
8/6/15 -- summary tables for a few energy and ToF cuts on January 2015 run data. Physics asymmetries and rates presented. Accompanying Asymmetry vs Thickness plots for the tables are also linked.
Media:Mott_thickness_study.ods
Time-of-Flight-Cut vs Not
2 sigma slices in energy from -3 to +3 sigma. Also -1 to +3 sigma and -3 to +3 sigma included. All raw data.
Media:2sigmaEslices_ToFCut.ods
Media:2sigmaEslices_noToFCut.ods
Sensitivity of A0 and Lambda to Systematic Time and Energy Cuts
Asymmetry versus thickness (with chi-squared values) plots for 1) FSEM measured thicknesses and uncertainties (same data as from July 14th presentation) 2)FSEM measured thicknesses and uncertainties adjusted to account for 5% variation within a given batch, as quoted be Lebow. 3)Lebow-quoted thicknesses (ie nominal thicknesses) with 10% uncertainty. 4) Table summarizing A_0s, lambdas and chi-squared values. To find χ2/ndf, divide the given chi-squared values by 10 (number of degrees of freedom is number of foils (12) minus number of fit parameters (2)).
1) Media:FSEM_uncertainty_only.odt
2) Media:FSEM_uncertainty_plus5.odt
3) Media:Lebow_quoted_thicknesses.odt
4) Media:systematics_table.odt
Summaries
Media:foil_Thickness_status-5June2015B.pptx
Media:foil_Thickness_status-20March2015.pptx
Media:foil thickness status_13Feb2015.pdf
Media:foil thickness status_13Feb2015.pptx
Media:first_thickness_meas.pptx
Working data
Media:Summary_all_foils_5June2015c.xlsx
Media:Elastic_rate_thickness_study_20March2015.xlsx
Media:Thickness_meas_summary_20March2015.xlsx
Media:Results_5613D_05_stat_analysis.xlsx
Media:Asy_vs_meas_thickness.xlsx
FE-SEM image data files
The output of the FE-SEM is a .tif file. The free software "ImageJ" has been used to measure the foil thickness from these images. The files with "meas" in the title have been output by the ImageJ software, and when they are opened again in that software, they show the lines/paths that were used to measure the thickness of the gold foils.
50 nm 6809B
media:Au_6809_B_50_cs_m03.tiff
media:Au_6809_B_50_cs_m04.tiff
media:Au_6809_B_50_cs_m06.tiff
media:Au_6809_B_50_cs_m03_meas.tiff
media:Au_6809_B_50_cs_m04_meas.tiff
media:Au_6809_B_50_cs_m06_meas.tiff
media:Au_6809_B_50_cs_m06_meas_even.tiff
media:Au_6809_B_50_cs_measurements.xlsx
225 nm 7029B
350 nm 5613A
media:Au_5613_D_350_cs_m01.tif
media:Au_5613_D_cs_m01_meas.tif
media:Au_5613_D_cs_m03_meas_even.tif
media:Au_5613_D_cs_m05_meas_even.tif
media:Au_5613_D_cs_m06_meas_even.tif
500 nm 5275C
Center
media:Au_5275_C_500_cntr_m05.tif
media:Au_5275_C_500_cntr_m06.tif
media:Au_5275_C_500_cntr_m07.tif
media:Au_5275_C_500_cntr_m07_meas.tif
Edge
media:Au_5275_C_500_edge_m01.tif
media:Au_5275_C_500_edge_m01_meas.tif
media:Au_5275_C_500_edge_m02.tif
media:Au_5275_C_500_edge_m02_meas.tif
media:Au_5275_C_500_edge_m05.tif
media:Au_5275_C_500_edge_m05_meas.tif
media:Au_5275_C_500_edge_m06.tif
media:Au_5275_C_500_edge_m06_meas.tif
media:Au_5275_C_500_edge_m07.tif
625 nm 7028B
media:Au_7028_B_625_m01_meas.tif
media:Au_7028_B_625_m03_meas.tif
media:Au_7028_B_625_m04_meas.tif
750 nm 5134B
media:Au_5134_B_750_cs2_m04.tif
media:Au_5134_B_750_cs2_m04_meas.tif
media:Au_5134_B_750_cs2_m05.tif
media:Au_5134_B_750_cs2_m05_meas.tif
890 nm 3057C
1000 nm 5385B